成果報告書詳細
管理番号20140000000172
タイトル*平成25年度中間年報 「次世代機能代替技術の研究開発 次世代再生医療技術の研究開発 生体内で自己組織の再生を促すセルフリー型再生デバイスの開発(幹細胞ニッチ制御による自己組織再生型心血管デバイスの実用化開発)」
公開日2014/8/8
報告書年度2013 - 2013
委託先名ニプロ株式会社
プロジェクト番号P10004
部署名バイオテクノロジー・医療技術部
和文要約
英文要約Title: Research and Development of Next-generation Regenerative Technology.
Development of cell-free regenerative device which promotes regeneration of autologous tissue in vivo.
Development of self-regenerating cardiovascular device by regulating stem cell niche.
Annual Report FY2013 (period: FY2010-FY2014).
(4) In vitro evaluation of device
We tried to combine stem cell inducing/differentiating factor with cardiac support device. 
We evaluated gelatin film in vivo and in vitro. One gelatin film including “stem cell inducing /differentiating factor”
is good in that it adheres closely to animal beating heart surface.The other film is cover layer to release slowly the
“stem cell inducing /differentiating factor” from the film mentioned above.
We succeeded to evaluate this new two-layer film system is good for slow-releasing “stem cell inducing /differentiating factor”
in vivo and in vitro.
(5) Optimizing a manufacturing process of device
We tried to get a best mode of optimizing manufacturing process of device using results of the past fiscal year.
We have done trials and errors in computer controlled knitting machine. Finally, we got the best mode of optimizing manufacturing process
about knitting program with repeatability. As a result, we established good and stable manufacturing process in a device size
and expansion. And we studied combination of gelatin film containing ONO1301- microsphere and cardiac support device.
And we got a new type of cardiac support device that had two pockets containing two gelatin films for posted both side of that one.
(7) Study of biological safety to cardiac support device
We performed intradermal reaction test as one of the biological safety tests of medical devices.
(8)Study of medical device sterilization method.
Current year, we studied device sterilization method. We found that ethylene oxide gas sterilization(EOG) will be suitable for
cardiac support device. There are three types of sterilization method in industrial medical device manufacture;1autoclaving,
2ethylene oxide gas sterilization(EOG), and, 3gamma-ray sterilization. Our device is made from biodegradable polymer fiber.
Autoclaving will destroy biodegradable polymer fiber through hydrolysis. EOG sterilization is best mode of sterilization method
in this cardiac support device, because of no influence on cardiac support device physical properties, and good effectiveness on sterilization.
(9)Market research
We are conducting market research. Current year, we do market research by using ‘’Statistical data of operating number pressed in Japanese.
ダウンロード成果報告書データベース(ユーザ登録必須)から、ダウンロードしてください。