タイトル*平成25年度中間年報 「次世代機能代替技術の研究開発 次世代心機能代替治療技術の研究開発 小柄患者用補助人工心臓の有効性及び安全性の評価(3)」
公開日2014/8/20
報告書年度2013 - 2013
委託先名独立行政法人国立循環器病研究センター
プロジェクト番号P10004
部署名バイオテクノロジー・医療技術部
和文要約
英文要約Evalution of effectiveness and safety of the implantable ventricular assist device for patients with small body sizes

Interim Report for FY 2013

Contractor : National Cerebral and Cardiovascular Center, Nipro Corporation, Mitsubishi Heavy Industries, Ltd.

Summary of the progresses

We worked on the search for an alternative permanent magnets of the axial flow pump, after discontinuation of the magnets from the supplier. We selected a set of substitute magnets that are theoretically equivalent to the former magnets. The first prototype containing the substitute magnets demonstrated to provide sufficient performance as designed. The prototype also contained a novel right-angle inlet port to miniaturize the whole dimension of the device. This right-angle port was expected to lower the pressure head due to its flow path, and its effect on the shear stresses and therefore on mechanical blood destruction should be investigated.
We carried out hemolysis testing in which the amount of broken red blood cells were measured to evaluate its safety in terms of blood losses with the device.
The hemolysis testing was carried out according to the international standard protocol using fresh calf blood in the laboratory of our facility. The measured amount of hemolysis caused by the prototype was within the normal range for clinical applications.