成果報告書詳細
管理番号20150000000609
タイトル*平成26年度中間年報 環境・医療分野の国際研究開発・実証プロジェクト 先進的医療機器システムの国際研究開発及び実証 現地国事情に適した高品位透析治療を達成する透析水清浄化システムの研究開発・実証(3)
公開日2015/8/13
報告書年度2014 - 2014
委託先名学校法人東京女子医科大学
プロジェクト番号P11011
部署名ロボット・機械システム部
和文要約
英文要約Title: International Research and Development and Demonstration Project of Environmental and Medical Field/ International Research and Development and Demonstration of Advanced Medical System/ Research and Development of Water Purification System for dialysate to Achieve High Quality Dialysis Treatment according to the affairs of each area (FY2012-FY2016)/ FY2014 Annual Report

In order to develop a safe and adequate system, we evaluated the effect of back-filtrated infusion on the solute removal in intermittent infusion hemodiafiltration (I-HDF) this year. Solute clearance of the hemodiafilter decreased with an increase in back-filtration flow rate, and its tendency was remarkable in larger molecules because their removal strongly depends on the convective transport due to back-filtration. However, its clinical impact would be negligible. To clarify the validity of the I-HDF, we would like to evaluate the effects of intermittent infusion pattern on solute clearance of the hemodiafilter and the improvement of peripheral circulation of the patient.
To establish a dialysis-dose monitor for uremic toxins contained in the spent dialysate, we also improved a prototype dialysis monitor using ultraviolet LED. We examined the performance of the prototype monitor using spent dialysate harvested by clinical hemodialysis treatments. Currently, we will also develop a navigation algorithm necessary to feedback to an optimal treatment conditions.
ダウンロード成果報告書データベース(ユーザ登録必須)から、ダウンロードしてください。